GS HIV-2 EIA

Print

Overview

GS HIV-2 EIA

GS HIV-2 EIA (for Detection of Ab to HIV-2)

Reproducibility of the HIV-2 EIA: Inter-Assay Reproducibility — Manual Testing

Specimen N Mean OD/CO SD %CV
1 36 4.927 1.010 20.5%
2 36 3.530 0.789 22.4%
3 36 2.799 0.897 32.0%
4 36 2.366 0.449 19.0%
5 36 1.821 0.401 22.0%
6 18 0.838 0.279 33.3%
7 16 0.205 0.056 27.3%
HIV-2 EIA Positive Control 276 1.228* 0.226 18.4%
HIV-2 EIA Negative Control 412 0.063* 0.018 28.6%


* Controls are expressed as mean OD values.

GS HIV-2 EIA Intra-Assay Reproducibility

GS HIV-2 EIA (for Detection of Ab to HIV-2)

Reproducibility of the HIV-2 EIA: Intra-Assay Reproducibility — Manual Testing

Specimen N Mean OD SD %CV
HIV-2 EIA Positive Control 288 0.900 0.040 4.4%
HIV-2 EIA Negative Control 288 0.060 0.008 13.3%
Positive Sample 288 0.585 0.028 4.8%
GS HIV-2 EIA Protocol

GS HIV-2 EIA (for Detection of Ab to HIV-2)

HIV-2 EIA Principle and Protocol

    
 
Pipette diluted sample into microwell. Incubate 1 hour at 37°C.
  
Wash microwell using wash solution.
  
Add conjugate reagent. Incubate 1 hour at 37°C.
  
Wash microwells. Add chromogen reagent. Incubate 30 minutes at room temperature.
  
Add stopping reagent. Read microplate at
450 nm.
 
HIV Serological Profile

HIV Serological Profile

The GS HIV-2 EIA is an enzyme immunoassay for the detection of circulating antibodies to human immunodeficiency virus type 2 (HIV-2) in human serum or plasma and is indicated as an aid in the diagnosis of infection with HIV-2.

The GS HIV-2 EIA is intended for manual use and also for use with the EVOLIS™ Automated Microplate System as an aid in the diagnosis of infection with HIV-2 or for follow-up testing for HIV-2 antibodies in blood donors.

GS HIV-2 EIA
GS HIV-2 EIA

32536
480 tests for detection of antibodies to human immunodeficiency virus type 2 (HIV-2) in human serum or plasma

List Price:   Inquire
Quantity:   Add to Quote