BioLogic LP System

BioLogic LP System

Overview

The BioLogic LP low-pressure chromatography system is used for biomolecule purification. Its compact design minimizes the workspace required in the coldroom or on the laboratory bench.
Read full description

Description

Description

The BioLogic LP low-pressure chromatography system is used for biomolecule purification. Its compact design minimizes the workspace required in the coldroom or on the laboratory bench. The system includes both 254 and 280 nm filters for nucleic acid and protein detection, and a conductivity cell to monitor gradient progress.

The system houses a peristaltic pump with a flow rate range of 0.05–40 ml/min (20 ml/min per channel; dual-channel peristaltic pump) and maximum backpressure of 30 psi (0.2 MPa); the system is compatible with Econo-Column low-pressure chromatography columns, UNOsphere, Bio-Gel P, and Macro-Prep chromatography media, Bio-Scale Mini and Econo-Pac Cartridges, Amersham HiTrap cartridges, SOURCE media, and all other low-pressure chromatography media. Addition of an SV-5 buffer select valve allows selection of up to 5 buffers and automated sample separation and loading of large sample volumes.

The system can collect into eighty 13 x 100 mm tubes or microtubes using the Model 2110 fraction collector, or collect into virtually any size container, from microplates to carboys, using the BioFrac fraction collector. In addition, the BioLogic LP system supports the use of many collectors from other manufacturers.

The BioLogic LP system has IEC 61010 safety certification.

LP Data View software captures data, multitasks, and prints data from any computer that runs on Windows 10 operating system. The system stores up to 50 methods; each method can include up to 50 pump steps and 50 fraction collection steps.

Specifications

Specifications

Pump
 
   Flow range
0.1–40 ml/min
   Maximim counterpressure
30 psi (2 bar)
   Gradient formation
Solenoid-activated proportioning valve
Mixer volume: 750 µl
Detection Performance
 
   UV
Single channel, 280/254 nm with auto zero
Detection range: 0.001–2 AUFS
   Conductivity
Detection range: 0.5–500 mS/cm FS
Temperature range
4–40°C
Sample Loading
Manual loop fill
Automatic loading through SV-5 buffer select valve
Technical Specifications
BioLogic LP System
Flow rate limit
20 ml/min per channel; dual-channel peristaltic pump
Pressure limit
30 psi (0.2 MPa)
UV Detection
254 and 280 nm filters; 2 mm flow cell
Conductivity
Standard
Buffer selection/large sample loading
Optional SV-5 buffer-select valve, Model EP-1 Econo pump or Econo gradient pump
Sample injection
Manual MV-6 valve
General fraction collection
BioFrac or Model 2110 fraction collector
Large-volume fraction collection
BioFrac fraction collector with optional prep adaptor
Large-volume fraction collection
BioFrac or Model 2110 fraction collector
Solvent Compatibility
All commonly used chromatographic solvents
Operating tempurature
4–40°C
Dimensions (W x D x H)
24 x 33 x 25 cm
(9.5 x 13 x 9.8")
Weight
19 kg (42 lb)

Ordering

Ordering

items

Use the filters below to refine results!

Image
731-8300_tn.jpg
7318300
100/120 V, low-pressure chromatography system, includes controller, rack, accessory kit, valve/mixer, UV optics, conductivity cell and holder, tubing and fittings, column holder, starter kit
Image
731-8302_tn.jpg
7318302
110/120 V, low-pressure chromatography system, includes standard BioLogic LP system (#731-8300), Model 2110 fraction collector (#731-8120), SV-3 diverter/bypass valve, system cable 1
Image
731-8338_tn.jpg
7318338
110/120 V, low-pressure chromatography system, includes standard BioLogic LP system, BioFrac fraction collector, LP Data View software, system cables 3 and 15, 25' serial cable
Image
731-8336_tn.jpg
7318336
110/120 V, low-pressure chromatography system, includes standard BioLogic LP system, Model 2110 fraction collector, LP Data View software, SV-3 diverter/bypass valve, system cable 1, 25' serial cable

Accessories

Accessories

Image
731-8320_tn.jpg
7318320
Pkg of 1, 6-port manual injection valve with female luer fittings, for use with BioLogic system rack and Econo gradient pump rack
Image
731-8321_tn.jpg
7318321
Pkg of 1, 5-port solenoid random-access valve, 4-position, 30 psi (2 bar) limit, for use with BioLogic LP systems and Econo gradient pumps
(Discontinued) Chart Paper
7318198
Pkg of 2, 10 cm x 13 m, chart paper for use with Model 1327 and 1325 chart recorders

Documents

Documents

Number Description Options
2038
BioLogic LP System Brochure, Ver C
4006073
Instruction Manual, BioLogic™ LP Starter Kit, Rev D