
Description
Description
Exact Diagnostics HIV-1 LOD Summary | |
Catalog Number | HIVLOD |
Analyte | Inactivated HIV-1 |
Matrix | True EDTA Plasma |
Preservative | ProClin® 300 |
Storage | -70°C or Below |
Fill Volume | 1.0 mL |
Number of Vials Per Kit | 10 Vials |
Number of Uses | Single Use |
Shelf Life | 24 Months from Date of Manufacture |
Traceability | 3rd WHO International Standard |
Precautions | Biological Risks |
Regulatory Status | In Vitro Diagnostic Use |
INTENDED USE:
The Exact Diagnostics HIV-1 LOD is calibrated against the 3rd WHO International Standard for HIV-1 (10/152)1. The Exact Diagnostics HIV-1 LOD is used to determine the analytical sensitivity of the assay for Human Immunodeficiency Virus (HIV-1) RNA. The Exact Diagnostics HIV-1 LOD allows laboratories to evaluate their molecular assay and test for operator proficiency.
PRODUCT DESCRIPTION:
The Exact Diagnostics HIV-1 LOD contains whole, intact virus that has been heat inactivated. The control is intended to test the entire process of a molecular assay including extraction, detection, and amplification.
The Exact Diagnostics HIV-1 LOD is formulated in True EDTA Plasma that is non-reactive for HBV DNA, HCV RNA, HIV-1 RNA, HBsAg and antibodies for HIV-1/2, and HCV.