Overview
| Exact Diagnostics HSV Verification Panel Summary | |
| Catalog Number | HSVP200 |
| Analyte 1 | Inactivated HSV-1 |
| Analyte 2 | Inactivated HSV-2 |
| Matrix | Synthetic Matrix |
| Preservative | Sodium Azide |
| Storage | -20°C or Below |
| Fill Volume | 1.0 mL |
| Number of Vials Per Kit | 9 Vials |
| Number of Uses | Single Use |
| Shelf Life | 18 Months from Date of Manufacture |
| Precautions | Biological Risks |
| Regulatory Status | In Vitro Diagnostic Use |
| Exact Diagnostics HSV Verification Panel Components |
| EDX HSV-1 3x LOD |
| EDX HSV-1 5x LOD |
| EDX HSV-1 10x LOD |
| EDX HSV-2 3x LOD |
| EDX HSV-2 5x LOD |
| EDX HSV-2 10x LOD |
| EDX HSV-1/HSV-2 3x LOD |
| EDX HSV-1/HSV-2 5x LOD |
| EDX HSV-1/HSV-2 10x LOD |
INTENDED USE:
The Exact Diagnostics HSV Verification Panel is intended to be used with HSV1 and HSV-2 assays to monitor the performance of that given assay. The Exact Diagnostics HSV Verification Panel is used to monitor the presence of Herpes Simplex Virus-1 (HSV-1) and Herpes Simplex Virus-2 (HSV-2) DNA. The Exact Diagnostics HSV Verification Panel allows laboratories to evaluate their molecular assay and test for operator proficiency.
PRODUCT DESCRIPTION:
The Exact Diagnostics HSV Verification Panel contains whole, intact virus that has been heat inactivated. EDX HSV Verification Panel is intended to test the entire process of a molecular assay including extraction, detection, and amplification. The Exact Diagnostics HSV Verification Panel is internally value assigned using Bio-Rad Digital Droplet PCR (ddPCR). The Exact Diagnostics HSV Verification Panel is formulated in Synthetic Matrix, which contains preservatives.