Description
Description
Exact Diagnostics EBV LOD Summary | |
Catalog Number | EBVLOD |
Analyte | Inactivated EBV |
Matrix | True EDTA Plasma |
Preservative | ProClin® 300 |
Storage | -70°C or Below |
Fill Volume | 1.0 mL |
Number of Vials Per Kit | 10 Vials |
Number of Uses | Single Use |
Shelf Life | 24 Months from Date of Manufacture |
Traceability | 1st WHO International Standard |
Precautions | Biological Risks |
Regulatory Status | In Vitro Diagnostic Use |
INTENDED USE:
The Exact Diagnostics EBV LOD is calibrated against the 1st WHO International Standard for Epstein-Barr Virus (09/260)1. The Exact Diagnostics EBV LOD is used to determine the analytical sensitivity of the assay for Epstein-Barr Virus (EBV) DNA. The Exact Diagnostics EBV LOD allows laboratories to evaluate their molecular assay and test for operator proficiency.
PRODUCT DESCRIPTION:
The Exact Diagnostics EBV LOD contains whole, intact virus that has been heat inactivated. The control is intended to test the entire process of a molecular assay including extraction, detection, and amplification. The Exact Diagnostics EBV LOD is formulated in True EDTA Plasma that is nonreactive for HBV DNA, HCV RNA, HIV-1 RNA, EBV DNA, HBsAg and antibodies for HIV-1/2, and HCV.