VARIANTnbs Newborn Screening System

VARIANTnbs Newborn Screening System

The VARIANTnbs System is the established worldwide standard in automated newborn screening for sickle cell disease and other hemoglobin disorders. 

Read More
Ordering Information

Overview

The VARIANTnbs System is designed for high-capacity screening and walk-away automation, reducing labor requirements and human error. Load up to three systems per workstation and up to 1,128 samples per run.

The VARIANTnbs System features fully automated analysis and advanced result "pattern" reporting for maximum efficiency. Using dried blood spot specimens, it detects hemoglobins (F, A, S, D, C, E) that are clinically significant for identifying:

  • Hemoglobin disorders  
  • Carrier status for abnormal hemoglobins 
  • Double heterozygous conditions 
  • Hemoglobin Barts flagged as FAB

 

The VARIANTnbs System provides the total hemoglobin picture

The system is efficient and easy to use. Smart reports with optional pattern rules simplify data review while providing objective, accurate results. A flexible worklist import function eliminates file incompatibility issues. Reviewing and managing results is simple with electronic chromatogram storage and retrieval. 

 

Educational Resources

Bio-Rad believes in sharing knowledge. We’ve created dedicated resources to help you strengthen your hemoglobinopathy expertise.

Image
Library of Variants Database screen

Library of Variants

In this digital database, find case reports to support the interpretation of hemoglobin variant chromatograms for sickle cell disease, thalassemia, and other hemoglobinopathies. Access this invaluable resource tool for your next case.

Image
Test Your Knowledge graphics

Test Your Knowledge

In this series of educational case studies, learn how to evaluate hemoglobinopathies and sickle cell cases and expand your knowledge of HPLC chromatograms.

Specifications

Dimensions
    VARIANTnbs System has two modules: Chromatography Station
    (VNCS) and Automatic Sampler (VNAS).Clearance for the system is: 
  • Left side of VNCS: 15 cm (6 in.)
  • Back: 10 cm (4 in.)
  • Right side of VNAS: 15 cm (6 in.)
  • Top: 5 cm above open instrument cover
  • Ensure the VNAS and VNCS are no more than 11.5 cm (4.5 in.) apart.

    Dimensions (H x W x D):
  • VNCS: 53.2 x 27.2 x 40.7 cm (20.9 x 10.7 x 16.0 in.)
  • VNAS: 26.4 x 47.0 x 44.0 cm (10.4 x 18.5 x 17.3 in.)
  • Waste Tank: 33.0 x 30.5 x 15.2 cm (13 x 12 x 6 in.)
    Weight (uncrated):
  • VNCS: ≤35 kg (77 lb)
  • VNAS: ≤20 kg (44 lb)
  • The bench or table should have a flat, level surface that is free from vibrations and is capable of supporting up to 90 kg (200 lb).
Electrical Data
Power Requirements: Input voltage AC 100-120 V/200-240 V, 50/60 Hz automatically selectable
    Power Consumption:
  • VNCS: 200 VA maximum (2 A maximum) = 682 BTU/hr
  • VNAS: 150 VA maximum (at Cooling ON) = 512 BTU/hr
Environment conditions
    Operating Environment:
  • Ambient temperature 15-25°C
  • Humidity 10-90%, non-condensing
    Storage Conditions:
  • Temperature -20°C to 50°C
  • Humidity 10-95%

Ordering

items

Use the filters below to refine results!

Image
nbs_250_3010_tn.jpg
250-3010

Instrument for newborn screening of hemoglobins S, F, A, C, D, and E, includes bar code reader for sample tracking