Overview
Liquichek Urine Toxicology Control, Levels S1O Low Opiate & S2O Low Opiate, are prepared from human urine containing drugs of abuse or their metabolites at concentrations 20–50% below or above commonly used screen cutoff levels.
- 2 year shelf life at -20°C to -70°C
- 30 day open-vial stability at 2–8°C
- 45 day closed vial stability at 2–8°C
- Includes Methamphetamine and Oxazepam
Refer to the package insert for performance and stability claims.
Gravimetric Reference Values
| Analytes | Level S1O Low Opiate (ng/mL) |
Level S2O Low Opiate (ng/mL) |
| Catalog Number | 683 | 706 |
| Amphetamines d-Amphetamine d-Methamphetamine MDMA |
– 750 – |
– 1,250 – |
| Barbiturates Secobarbital |
150 |
375 |
| Benzodiazepines Lormetazepam Nordiazepam Oxazepam |
– – 100 |
– – 300 |
| Cannabinoids 11-nor-Δ-9-THC-9-COOH |
37.5 |
65 |
| Cocaine Benzoylecgonine |
225 |
375 |
| Ethanol (mg/dL) | 40 | 70 |
| Lysergic Acid Diethylamide (LSD) | 0.3 | 0.8 |
| Methadone | 225 | 375 |
| Methaqualone | 225 | 375 |
| Opiates Morphine (Free) |
225 |
375 |
| Phencyclidine (PCP) | 19 | 31 |
| Propoxyphene | 225 | 375 |
Ordering
Image
(Discontinued) Liquichek Urine Toxicology Control, Level S1O Low Opiate
683
Liquid human urine screening control with drugs of abuse at 20 - 50% below commonly used cutoff levels; for 300 ng/mL Opiate cutoff (10 x 10 mL)
Image
(Discontinued) Liquichek Urine Toxicology Control Level S2O Low Opiate
706
Liquid human urine screening control with drugs of abuse at 20 - 50% above commonly used cutoff levels; for 300 ng/mL Opiate cutoff (10 x 10 mL)
Image
683X
Liquid human urine screening control with drugs of abuse at 20 - 50% below commonly used cutoff levels; for 300 ng/mL Opiate cutoff; MiniPak (1 x 10 mL)
Image
706X
Liquid human urine screening control with drugs of abuse at 20 - 50% above commonly used cutoff levels; for 300 ng/mL Opiate cutoff; MiniPak (1 x 10 mL)
Image
(Discontinued) Liquichek Urine Toxicology Negative Control
460
Liquid human urine negative control confirmed negative at the specified detection limits for over 30 drugs of abuse (10 x 20 mL)
Image
(Discontinued) Liquichek Urine Toxicology Negative Control, MiniPak