PrimePCR™ Probe Assay: TIMP3, Human

RT

Real-time PCR probe assay designed for gene expression analysis. Probe assays consist of unlabeled PCR primers and a dual labeled fluorescent probe.

List Price:    $202.00
Your Price:   Log In
 

Assay Information

Technology:   qPCR
Assay Type:   Probe
Application:   Gene Expression
Unique Assay ID:   qHsaCIP0029799
Assay Design:   Intron-spanning
Chromosome Location:   22:33245484-33253346question
Amplicon Length:   119
Splice Variants Targeted:   ENST00000266085 ENST00000382049 ENST00000538671

Gene Information

This gene belongs to the TIMP gene family. The proteins encoded by this gene family are inhibitors of the matrix metalloproteinases a group of peptidases involved in degradation of the extracellular matrix (ECM). Expression of this gene is induced in response to mitogenic stimulation and this netrin domain-containing protein is localized to the ECM. Mutations in this gene have been associated with the autosomal dominant disorder Sorsby's fundus dystrophy. [provided by RefSeq Jul 2008]

Gene Symbol:   TIMP3
Gene Name:   TIMP metallopeptidase inhibitor 3
Aliases:   HSMRK222, K222, K222TA2, SFD
RefSeq:   NC_000022.10 NG_009117.1 NT_011520.12
Ensembl:   ENSG00000100234
Entrez:   7078
UniGene:   Hs.644633
Chromosome Mapping:   22q12.1-q13.2,22q12.3

The below validation information is for the Primer Pair only   Download Validation Data (.pdf)


Products used to generate validation data:

Real-Time PCR Instrument CFX384 Real-Time PCR Detection System
Reverse Transcription Reagent iScript™ Advanced cDNA Synthesis Kit for RT-qPCR
Real-Time PCR Supermix SsoAdvanced™ SYBR® Green Supermix
Experimental Sample qPCR Human Reference Total RNA

Summary Data:

R2 0.999400
y-intercept 34.390000
Efficiency 102

Amplification Plot
Amplification of cDNA generated from universal RNA.

Amplification of cDNA generated from universal RNA.

Melt Peak
Melt curve analysis of above amplification.

Melt curve analysis of above amplification.

Standard Curve
Standard curve generated using 20 million copies of template diluted 10 fold to 20 copies.

Standard curve generated using 20 million copies of template diluted 10 fold to 20 copies.

Number Description Download