Bio-Plex Pro™ Mouse Th17 Assay Specifications
| |
Target |
LLOQ |
ULOQ |
LOD |
Intra-Assay %CV |
Inter-Assay %CV |
| 8-Plex |
IL-17F |
1 |
6,362 |
0.1 |
5 |
5 |
| IL-21 |
6 |
8,207 |
2.8 |
3 |
15 |
| IL-22 |
0.5 |
8,433 |
0.1 |
4 |
5 |
| IL-23p19 |
3.4 |
55,764 |
0.7 |
5 |
6 |
| IL-31 |
2.5 |
41,313 |
5.2 |
6 |
6 |
| IL-33 |
0.5 |
7,343 |
0.2 |
5 |
12 |
| CD40L |
3.5 |
11,245 |
0.4 |
4 |
14 |
| MIP-3α |
0.7 |
1,660 |
0.3 |
6 |
13 |
| |
| 6-Plex |
IL-1α |
10.36 |
60,631 |
9.4 |
4 |
7 |
| IL-6 |
0.74 |
12,053 |
0.2 |
3 |
16 |
| IL-10 |
2.95 |
12,066 |
1 |
4 |
5 |
| IL-17A |
2.65 |
43,337 |
0.8 |
3 |
10 |
| IFN-γ |
1.84 |
30,164 |
1.2 |
4 |
6 |
| TNF-α |
5.8 |
59,626 |
1.4 |
3 |
6 |
| |
| 2-Plex |
IL-25 |
168 |
98,266 |
9 |
9 |
12 |
| IL-27p28 |
63.4 |
117,521 |
29.2 |
9 |
6 |
| |
| Singleplex |
ICAM-1 |
2.9 |
23,723 |
0.78 |
4 |
8 |
Assay Performance Definitions
Assay working range — the range of concentrations within which the assay is precise and accurate. Boundaries of the assay working range are the lower limit of quantitation (LLOQ) and the upper limit of quantitation (ULOQ).
Precision — the coefficient of variation (%CV) at concentrations within the assay working range
Accuracy — ratio of the observed concentration versus the expected concentration of a known amount of spiked analyte within the assay working range
Sensitivity (limit of detection, LOD) — the concentration of analyte for which the fluorescence intensity signal is 2-fold of the standard deviation above the background signal