Profinia Software and Firmware Updates
Profinia Software and Firmware
If you purchased the Profinia software and firmware after June 10, 2008, your system will have updated versions preinstalled. If your system was bought before this date follow the directions below.
Profinia Software
Version 2.0 is available to replace Version 1.0 as of June 10, 2008. Call Bio-Rad Technical Support (1-800-424-6723, option 2) to get the updated version.
Profinia Firmware
Latest firmware — main board v1.40; pump and valves (PV) v2.01, are available for download.
For best performance, use new (brown) inlet check valves (#620-0413) with PV firmware 2.01.
To check the firmware version of the Profinia system, go to Data/Utilities on the home screen, select Maintenance/Functions, and then Display Firmware Versions. For proper use of Protein A or Profinity eXact purification methods, the main board firmware should be updated to v1.40.
Downloads If you have Profinia software version 2.00, click here to upgrade to version 2.01.
Click here for main board (UI) v1.40 upgrade.