Model Characteristics |
| |
|
PharosFX Plus |
PharosFX |
PMI |
| Detection limit |
| Storage phosphor |
<0.95 dpm/mm2 for one exposure to 14C using imaging screen-K |
X |
|
X |
| <0.15 dpm/mm2 for one exposure to 32P using imaging screen-K |
X |
|
X |
Fluorescence (depends on experimental conditions) |
0.2 fmol of FITC end-labeled DNA using 488 nm laser |
X |
X |
|
| 6 pg of SYBR® Green I-stained DNA using 488 nm laser |
X |
X |
|
| 0.4 fmol of FITC end-labeled DNA using 532 nm laser |
X |
X |
|
| 25 pg of SYBR® Green I-stained DNA using 532 nm laser |
X |
X |
|
| 0.2 fmol of Cy3 end-labeled DNA using 532 nm laser |
X |
X |
|
| 0.2 fmol Cy5 end-labeled DNA using 635 nm laser |
X |
X |
|
| Dynamic Range |
5 orders of magnitude (1:65,535) or 16-bit |
X |
X |
X |
| Linearity |
r2 > 0.99 |
X |
X |
X |
| Uniformity |
±5% over entire scan area |
X |
X |
X |
| Scan resolution |
800, 200, 200 and 50 µm (user selectable) |
X |
X |
X |
| Scan time |
20 x 25 cm area; 8.5 min at 100 µm, 15 min at 50 µm |
X |
X |
X |
| 35 x 43 cm area; 8.5 min at 200 µm, 17 min at 100 µm |
X |
X |
X |
| Spatial resolution of storage phosphor* |
14C:200 µm (2.5 line pairs/mm) using imaging screen-K |
X |
|
X |
| 32P:300 µm (1.5 line pairs/mm) using imaging screen-K |
X |
|
X |
| Digital resolution |
16-bit (65,636 gray scale) |
X |
X |
X |
| Excitation source |
25 mW 532 nm (green) diode-pumped solid-state laser |
X |
X |
|
| 10 mW 635 nm diode laser |
X |
|
X |
| Optional external lasers |
15 mW 488 nm (blue) extermal argon ion laser |
X |
X |
|
| 10 mW 635 nm (red) extermal diode laser |
X |
X |
|
| Maximum power |
65 W |
X |
X |
|
| Input voltage range |
100–240 VAC, 50–60 Hz |
X |
X |
X |
| Operating humidity requirements |
10–32°C, 30–80% |
X |
X |
X |
| Computer interface |
USB2 |
X |
X |
X |
| Operating system |
Windows 2000, XP, Vista or Mac OS X |
X |
X |
X |
Dimensions (W x D X H) |
57 x 66 x 30 cm (22.4 x 26.8 x 11.8") |
X |
X |
X |
| Weight |
32 kg (70.5 lb; scanner) |
X |
X |
X |