Wavelength
|
254 and 280 nm
|
|
Sensitivity ranges
|
2.0, 1.0, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01 AUFS
|
|
Detection limit
|
7 µg/ml (BSA in H2O)
|
|
Lamp
|
Low-pressure mercury with phosphor screen
|
|
Filters
|
254 and 280 nm
|
|
Output signal
|
0–1 V analog (impedance 150 Ω)
|
|
Operating temperature
|
4–40°C
|
|
Flow cell
|
Optical path 2 mm, internal volume 80 µl, illuminated volume 3 µl
|
|
Noise
|
1 x 10–4 OD max, peak-to-peak (dry cell); 2 x 10–4 OD max, peak-to-peak (flowing liquid)
|
|
Safety
|
Meets IEC 61010 and CSA 22.2 certification
|
|
Dimensions (W x D x H)
|
Base unit: 14.6 x 18.6 x 20.2 cm Optics unit: 13.2 x 15.2 x 3.8 cm
|