Environment conditions
|
|
|
Operating temperature
|
15–30°C (59–86°F)
|
|
Operating humidity, noncondensing
|
20–80%
|
|
Operating altitude
|
Designed to operate at up to 2,000 m (6,600') mean sea level
|
|
Compensatory range
|
±2°C
|
|
UL installation category
|
UL Installation Category II, as defined in Annex J of UL 3101–1
|
|
Pollution degree
|
UL Pollution Degree 2, as defined in Section 3.7, 3.2, of UL 3101–1
|
|
Bio-Plex Array Reader Specifications
|
|
|
Input voltage range
|
100–240 V, 1.5 A, 47 – 63 Hz
|
|
Physical dimensions (W x D x H), cm
|
43 x 51 x 23
|
|
Weight, kg/lb
|
23/60
|
|
Lasers
|
|
|
Reporter laser
|
532 nm, >10 mW, frequency-doubled diode; 30 x 60 µm elliptical beam
|
|
Classification laser
|
635 nm, 10 mW, diode; 30 x 60 µm elliptical beam
|
|
Fluidics
|
|
|
Sheath flow rate, µl/sec
|
90
|
|
Cuvette
|
200 µm square flow channel
|
|
Sample injection rates, µl/sec
|
60
|
|
Electronics
|
|
|
Reporter channel detection
|
Photomultiplier tube, A/D resolution 14 bits
|
|
Classification and doublet discriminator channel detection
|
Avalanche photodiodes with temperature compensation, A/D resolution 12 bits
|
|
Communication interface
|
RS–232
|
|
Signal Processing
|
|
|
Measurement resolution
|
15 bits effective
|
|
Processor modes
|
Linear, with logarithmic or linear display option
|
|
Dynamic range, dB
|
70
|
|
Bio-Plex Microplate Platform Specifications
|
|
|
Input voltage range
|
100–240 V, 1.8 A, 47–63 Hz
|
|
Physical dimensions (W x D x H), cm
|
44 x 61 x 8
|
|
Weight, kg/lb
|
14.4 (32)
|
|
Communications interface
|
RS–232
|
|
Plate capacity
|
One 96-well microtiter plate, ≤0.75
|