Refer to the PR3100 TSC User Manual and PR3100 TSC Microplate Assay Testing Reference Manual (MATRM) for more detailed instructions.
| 2. Running the QC Pac2 |
| a. Touch [Q2] box on the top toolbar. |
| b. Follow instruction on screen. Touch [»]. |
| c. Enter User Name. |
d. Enter QC Plate #. - Enter #, first time the plate is used.
- Touch [⇓] to show previous Plate # to be selected.
|
| e. Place the Verification Plate on the plate support with marking A1 on the plate in the top left corner. |
| f. Touch [»]. Reader will start reading the plate. |
| g. When the plate comes out and stays out, turn the plate 180° with marking A1 at the bottom right corner. |
| h. Touch [»]. Reader will start reading the plate. |
| i. When the plate comes out and stays out for the second time, take the plate off the plate support and place it back into the plastic bag before putting back into the case. |
| j. Touch [Move…] box to access plate movement. |
| k. Retract Plate Support by touching [In]. |
| l. Touch [Finish] to go to save result option — default filename contains the Date in the filename. |
| m. Touch [OK] and Save. |
| n. Review screen for any number other than “0” next to FAIL. “0” means all passed. |
| o. Touch [Printer Icon] to print the complete report. Select the printer driver. Default is PCL Compatible Printer (Color). |
| p. Touch [OK] to print complete report. |
| q. Touch File on Toolbar followed with Exit. |
| r. Power down Reader with switch on the rear right. |
| s. Take out the QC Pac2 USB stick and replace with the APF USB stick. |
| t. Ready for Daily Start-up. |