Comparison with EIA Methods
Positive Agreement
A total of 227 retrospective samples positive for anti-MPO (N = 100), anti-PR3 (N = 100) and anti-GBM (N = 27) were testedwith the BioPlex™ 2200 Vasculitis Kit and the corresponding commercially available microplate EIA methods. Thepositive % agreement and overall % agreement between the two methods for each analyte are referenced in thetable below. The BioPlex 2200 Vasculitis Kit demonstrated good positive agreement with commerciallyavailable EIA assays for each analyte.
| | BioPlex 2200 Vasculitis Anti-MPO Result | BioPlex 2200 Vasculitis Anti-PR3 Result | BioPlex 2200 Vasculitis Anti-GBM Result | | | Positive % Agreement | Overall % Agreement | Positive % Agreement | Overall % Agreement | Positive % Agreement | Overall % Agreement | EIA Result Retrospective Samples Positive for Anti-MPO, Anti-PR3 and Anti-GBM | 93.9% (92/98) | 93.0% (93/100)* | 100% (79/79) | 83.0% (83/100)† | 88.9% (16/18)‡ | 92.6% (25/27) | |
* One (1) anti-MPO EIA equivocal result is included in the Overall Agreement. † Ten (10 anti-PR3 EIA equivocal results are included in the Overall Agreement. ‡ Two (2) of the sixteen (16) BioPlex 2200 anti-GBM positive results were weak positives by anti-GBM EIA. ‡ One (1) of the two (2) BioPlex 2200 anti-GBM negative results was a weak positive by anti-GBM EIA. |
Negative Agreement
A total of 293 samples from normal blood donors and 300 samples from unselected patient samples previously tested with vasculitis tests were tested with the BioPlex 2200 Vasculitis kit and the corresponding commercially available microplate EIA methods. The negative % agreement and overall % agreement between the two methods for each analyte are referenced in the table below. The BioPlex 2200 Vasculitis kit demonstrated excellent negative agreement with commercially available EIA assays for each analyte.
| | BioPlex 2200 Vasculitis Anti-MPO Result | BioPlex 2200 Vasculitis Anti-PR3 Result | BioPlex 2200 Vasculitis Anti-GBM Result |
| | Negative % Agreement | Overall % Agreement | Negative % Agreement | Overall % Agreement | Negative % Agreement | Overall % Agreement |
EIA Result Normal Blood Donors | 100% (293/293) | 100% (293/293) | 100% (293/293) | 100% (293/293) | 93.3% (289/291) | 98.6% (289/293) |
EIA Result Unselected Patient Samples* | 97.6% (284/291) | 96.3% (289/300) | 99.0% (292/295) | 99.0% (297/300) | 99.7% (298/299) | 99.3% (298/300) |
* Previously tested with vasculitis tests.
Positive Agreement
A total of 200 retrospective positive samples (100 anti-MPO and 100 anti-PR3) were evaluated by a commercially available ANCA IFA method using ethanol-fixed slides. The positive agreement and overall agreement between the two methods are referenced in the tables below. The BioPlex 2200 Vasculitis Kit demonstrated good positive agreement with the ANCA IFA method.
| | BioPlex 2200 Anti-MPO Result |
| | Positive % Agreement | Overall % Agreement |
pANCA IFA Result Retrospective Samples Positive for Anti-MPO | 93.3% (83/89) | 84.0% (84/100) |
| | BioPlex 2200 Anti-PR3 Result |
| | Positive % Agreement | Overall % Agreement |
pANCA IFA Result Retrospective Samples Positive for Anti-PR3 | 94.9% (93/98) | 93.0% (93/100) |
BioPlex 2200 Vasculitis vs. EIA Methods
For the 200 retrospective positive samples described above, a combined positive agreement of 94.1% was observed between BioPlex 2200 Vasculitis anti-MPO/anti-PR3 and ANCA IFA results, compared to a combined positive agreement of 88.2% between anti-MPO/anti-PR3 EIA and ANCA IFA results. Also, a combined overall agreement with ANCA IFA results of 88.5% and 83.0% was observed between BioPlex 2200 and EIA results, respectively. Overall, the BioPlex 2200 Vasculitis Kit demonstrated better concordance with ANCA IFA than the corresponding EIA methods.
| | Combined BioPlex 2200 Anti-MPO/Anti-PR3 Results | Combined Anti-MPO/Anti-PR3 EIA Results |
| | Positive % Agreement | Overall % Agreement | Positive % Agreement | Overall % Agreement |
Combined p-/c-ANCA IFA Results | 94.1% (176/187) | 88.5% (177/200) | 88.2% (165/187) | 83.0% (166/200) |