| Specification |
Description |
| Sample capacity |
Standard tray: 84 vials of 1.5 ml and 3 vials of 10 ml (standard) |
| Vial dimension (cap included) |
Maximum vial height: 47 mm Minimum vial height: 32 mm |
| Loop volume |
2,000 µl (standard) or 50, 100, 200, 250, 500, 1,000, and 10,000 µl (optional) loops |
| Dispenser syringe |
1,000 µl (standard) or 250 or 2,500 µl (optional) syringes |
| Missing vial detection |
Missing vial detection by vial sensor |
| Headspace pressure |
Built-in compressor |
Switching time injection valve |
Electrically, <100 ms |
| Needles |
0.50 mm ID silicon-coated needle and air sleeve (standard) or analytical syringe and airsleeve (optional with analytical kit) |
| Piercing precision needle |
±0.6 mm |
| Wash solvent |
External wash solvent bottle |
| Wetted parts in the flow path |
SS 316, PTFE, TEFZEL, VESPEL, glass, Teflon, PEEK |
| Prep options |
Large-volume tray (24 x 10 ml vials), 2.5 ml syringe, 0.50 mm ID needle, large-bore PEEK injection valve, 10 ml sample loop |
| Analytical options |
250 µl syringe, analytical needle, 100 and 200 µl sample loop |
| Bio-compatible valve |
Special PEEK injection valve, quick-connect mounting |
| Sample tray cooling |
Built-in Peltier cooling (optional installed): • Range: 4–15°C • Accuracy: ±2°C (at temperature sensor) • Temperature at relative humidity of 80% and ambient temperature of 30°C |
| Cooling capacity |
20°C below ambient |
| Methods |
9 programmable methods |
| Injection method |
Injection mode (standard configuration): • Full loop injection • Partial loop fill injection • µl pick-up injection
|
| Injection volume |
5–1,500 µl (standard configuration), max 5,000 µl with Prep Bio kit |
| Injections per vial |
Max 9 injections |
| Wash |
Programmable: • 4 x AUX-1 ON/OFF • 4 x AUX-2 ON/OFF |
| Series |
9 programmable series |
| Priority sample |
Programmable |
| Reproducibility |
• RSD ≤0.3% for full loop injections • RSD = 0.5% for partial loop fill injections, injection volumes >10 µl • RSD ≤1.0% for µl pick-up injections, injection volumes >10 µl
|
| Memory effect |
<0.05% with programmable needle wash |