Operating mode
|
Conductivity, in Siemens (S)
|
|
Operating temperature
|
4–40°C
|
|
Sensitivity ranges (full scale)
|
0–10 µS, 0–100 µS, 0–10 mS,
|
|
Output signal
|
0–1 V
|
|
Power requirements
|
90–264 V, 50–60 Hz (depending upon power adaptor used)
|
|
Power consumption
|
20 W maximum
|
|
Noise
|
Short term: <0.1 µS Long term: <0.5 µS
|
|
Linearity
|
±0.5% full scale
|
|
Accuracy
|
±2% full scale
|
|
Drift
|
<0.5 µS/hr/°C
|
|
Dimensions
|
Control unit (W x L x H): 15.60 x 22.86 x 8.89 cm Flow cell (diameter x L): 3.1 x 3.8 cm
|
|
Weight
|
Control unit: 600 g
|
|
Case material
|
Chemical and fire resistant polypropylene
|
|
Regulatory
|
CSA: C22.2 No. 151-M1986
|
|
Compliance
|
TUV: EN61010–1:1993 CE: EN50082–2:1995 EN55011: 1991 Class A
|
|
Flow Cell
|
|
|
Type
|
Flow through, externally mounted
|
|
Temperature compensation
|
0–60 µC
|
|
Internal volume
|
8 µl, swept volume
|
|
Cell constant
|
Approximately 40°
|
|
Maximum operating pressure
|
60 psi, 4 bar, or 0.41 mPa
|
|
Cable length
|
1.22 m
|
|
Wetted parts
|
Titanium, Kel-F, and Tefzel
|
|
Fittings
|
Removable luer (will also accept 1/4"–28 flat bottom fittings)
|