Correlation
Correlation with CDC Evaluation Serum Panels
The Centers for Disease Control (CDC) provided evaluation serum panels for testing of T. gondii IgG, Rubella IgG and CMV IgG. These panels were tested to evaluate performance of the ToRC IgG kit and do not imply an endorsement of the BioPlex™ 2200 ToRC IgG kit by the CDC.
Table 1. BioPlex 2200 Results for CDC T. gondii. Reference Sera (n=100)
| CDC Samples | Reference Result | BioPlex Pos (+) | BioPlex Neg (–) |
| T. gondii | Pos (+) | 70 | 0 |
| Neg (–) | 0 | 30 |
Note: Tested at Bio-Rad Laboratories.
Table 2. BioPlex 2200 Results for CDC Rubella IgG Reference Sera (n=100)
| CDC Samples | Reference Result | Site 1 | Site 2 | Site 3 |
| BioPlex Pos (+) | BioPlex Neg (–) | BioPlex Pos (+) | BioPlex Neg (–) | BioPlex Pos (+) | BioPlex Neg (–) |
| Rubella IgG | Pos (+) | 82 | 0 | 82 | 0 | 82 | 0 |
| Neg (–) | 0 | 18 | 0 | 18 | 0 | 18 |
Note: Tested at 3 clinical evaluation sites.
Table 3. BioPlex 2200 Results for CDC T. gondii Reference Sera (n=100)
| CDC Samples | Reference Result | BioPlex Pos (+) | BioPlex Neg (–) |
| CMV IgG | Pos (+) | 66 | 0 |
| Neg (–) | 0 | 34 |
Note: Tested at Bio-Rad Laboratories