Components of a vertical electrophoresis cell.
Vertical electrophoresis cells are manufactured in different size formats to accommodate different gels sizes. Cell selection is dependent on the requirements for speed, resolution, and throughput (both number of samples and number of gels) as well as the volume of sample available (see Table below).
Vertical Electrophoresis System Selection Guide
Vertical Electrophoresis System Selection Guide
- Mini-format systems — accommodate small gels (up to 8.6 x 6.7 cm). The short separation distance maximizes the electrical field strength (V/cm) to yield rapid separations with moderate resolution. Use these systems for rapid analysis, method development, or when sample volumes are limited. The Mini-PROTEAN® system includes the Mini-PROTEAN Tetra cell (with a capacity of up to four gels) and the high-throughput Mini-PROTEAN® 3 Dodeca™ cell (for running up to 12 gels)
- Midi-format systems — accommodate 13.3 x 8.7 cm gels and offer rapid runs with higher sample capacity and enhanced separation over mini-format gels. The Criterion™ system includes the Criterion™ cell (for 1–2 gels) and high-throughput Criterion™ Dodeca™ cell (for 1–12 gels); both cells are compatible with Criterion precast gels
- Large-format systems — accommodate large gels (up to 20.0 x 18.3 cm for the PROTEAN® II system, designed for one-dimensional PAGE) and so offer maximum resolution. The PROTEAN II system provides a choice of glass plates, spacer, and sandwich clamps to cast two gel lengths/ 10 or 20 cm. The PROTEAN® Plus Dodeca™ cell allows maximum throughput with the capability to run up to 12 2-D gels at a time
Power supplies are available to meet the power requirements of numerous applications (see below). The choice of power supply for PAGE applications usually depends on the size and number of gels being run. The following Table compares the Bio-Rad PowerPac™ power supplies recommended for vertical electrophoresis applications.
- For mini-format vertical PAGE applications, the PowerPac Basic and PowerPac HC power supply are both suitable
- For large-format vertical PAGE applications, the PowerPac HV and PowerPac Universal power supply are both suitable
- For applications that require high currents, such as PAGE with the high-throughput Dodeca cells, the PowerPac HC power supply is the appropriate model
PowerPac power supplies selection guide
| Technique and Recommended Apparatus |
Typical Conditions* (Initial) |
Typical Conditions* (Final) |
Run Time |
PowerPac™ Power Supply |
| V |
mA |
V |
mA |
| Laemmli (SDS), O'Farrell Second Dimension (SDS) |
| PROTEAN® II xi cell |
100 |
35(C) |
350 |
35(C) |
5 hr |
HV or Universal |
| PROTEAN II XL cell |
100 |
35(C) |
350 |
35(C) |
5 hr |
HV or Universal |
| Criterion™ cell |
200(C) |
200 |
200(C) |
80 |
55 min |
Basic or HC |
| Mini-PROTEAN® Tetra cell |
200(C) |
240 |
200(C) |
120 |
35–45 min |
Basic or HC |
| High-Throughput Electrophoresis |
| PROTEAN Plus |
200(C) |
1,000 |
200(C) |
350 |
6 hr |
HC or Universal |
| Dodeca™ cell |
200(C) |
1,000 |
200(C) |
350 |
6 hr |
HC or Universal |
| 150(C) |
1,200 |
150(C) |
300 |
18–20 hr |
HC or Universal |
| PROTEAN II xi/XL multi-cell |
150 |
480(C) |
500 |
480(C) |
5 hr |
Universal |
| Criterion™ Dodeca™ cell |
200(C) |
1,000–1,400 |
200(C) |
400–500 |
55 min |
HC or Universal |
| Mini-PROTEAN® 3 Dodeca™ cell |
200(C) |
600 |
200(C) |
360 |
45 min |
HC or Universal |
| Western Blotting |
| Mini Trans-Blot® cell |
100(C) |
250 |
100(C) |
450 |
1 hr |
HC |
| Criterion™ blotter |
| Wire electrodes |
100(C) |
250 |
100(C) |
450 |
1 hr |
HC |
| Plate electrodes |
100(C) |
650 |
100(C) |
1,600 |
30 min |
HC |
| Trans-Blot® cell |
| Wire electrodes |
60(C) |
210–250 |
60(C) |
210—250 |
5 hr |
HC |
| Plate electrodes |
100–150(C) |
1,000–1,600 |
100–150(C) |
1,000–1,600 |
1–5 hr |
HC |
| High-intensity transfer |
50–100(C) |
1,600 |
50–100(C) |
1,600 |
30 min |
HC |
| Trans-Blot® Plus cell |
100(C) |
3,000 |
100(C) |
3,000 |
30 min–1 hr |
HC |
| Trans-Blot® SD cell |
| Protein |
15(C) |
500 |
15(C) |
200 |
15–30 min |
HC |
| DNA/RNA |
15 |
650(C) |
25 |
650(C) |
10–30 min |
HC |
*(C) = constant. Typical conditions are listed as guidelines only and can vary based on sample, buffers, etc.