S3™ Cell Sorter (488 nm) with S3 Biosafety System Class I #145-1021

Print

Power requirements 100–240 VAC, 50–60 Hz, 1.25 Amp
Power and efficiency 125 W
Heat load 50 W at <85 BTU/hr
Airflow 175 CFM (cubic feet per minute)
Blower R1G 175 DC motorized impeller
Fuse 2.0 A slow blow
Weight  52 lb
Dimensions (W x H x D), cm 135 x 146 x 96 cm, 53 x 57.5 x 37.75 in
Sound level 65 dB at maximum airflow
Operating temperature 18–25°C
Operating humidity 20–60%
HEPA filtration 99.997% or better on 0.3 μm- IEST-RP-CC-034
HEPA filters 23.625 x 23.625 x 3.54 in, certify annually
Top panel Clear polycarbonate
Base floor High-density polyethylene
Vinyl sides 30 mil PVC film and 18 oz vinyl-coated polyester
Frame 25 mm anodized aluminum T-slot profile

Use the S3 cell sorter with the S3 biosafety system Class I containment hood to sort cells expressing fluorescent proteins or with fluorescent markers such as labeled antibodies. This biosafety system protects both users and the environment from aerosols created during the cell sorting process.

This compact, affordable, integrated cell sorting system includes the following:

  • S3 cell sorter with 1 laser (488 nm) and 2 detectors (145-1001)
  • S3 biosafety system Class I (145-1078)

Related Products

The S3 cell sorter is available in 1-laser, 2-detector and 2-laser, 4-detector versions either alone or bundled with the S3 biosafety system Class I.

  • S3 cell sorter with 2 lasers (488/561 nm) and 4 detectors (145-1002)
  • S3 cell sorter with 2 lasers (488/561 nm), 4 detectors, and the S3 biosafety system Class I (145-1022)