ToRC IgG

Print

Overview

BioPlex™2200 ToRC IgG

Overview

Benefits of Multiplex Testing


BioPlex™2200 ToRC IgG

Reproducibility

  With-Run CV Between-Run %CV Total Precision %CV
T. gondii 4.6–7.4 0.0–2.0 6.0–14.6
Rubella 4.4–6.6 0.0–2.1 6.5–10.1
CMV 2.7–5.8 0.0–3.4 6.4–11.3
BioPlex™ 2200 ToRC IgG

ToRC IgG Assay Procedure


1. Patient sample (5 µL) automatically added to
    reaction vessel

2. Sample diluent & beads (dilution 1:60) added
    and incubated at 37°C

3. Wash step

4. Conjugate added and incubated at 37°C

5. Wash step

6. Bead resuspended in wash buffer

7. Flow-based, dual laser detection

BioPlex™2200 ToRC IgG

Correlation

Correlation with CDC Evaluation Serum Panels

The Centers for Disease Control (CDC) provided evaluation serum panels for testing of T. gondii IgG, Rubella IgG and CMV IgG. These panels were tested to evaluate performance of the ToRC IgG kit and do not imply an endorsement of the BioPlex™ 2200 ToRC IgG kit by the CDC.

Table 1. BioPlex 2200 Results for CDC T. gondii. Reference Sera (n=100)

CDC Samples Reference Result BioPlex Pos (+) BioPlex Neg (–)
T. gondii Pos (+) 70 0
Neg (–) 0 30


Note: Tested at Bio-Rad Laboratories
.


Table 2. BioPlex 2200 Results for CDC Rubella IgG Reference Sera (n=100)

CDC Samples Reference Result Site 1 Site 2 Site 3
BioPlex Pos (+) BioPlex Neg (–) BioPlex Pos (+) BioPlex Neg (–) BioPlex Pos (+) BioPlex Neg (–)
Rubella IgG Pos (+) 82 0 82 0 82 0
Neg (–) 0 18 0 18 0 18


Note: Tested at 3 clinical evaluation sites
.


Table 3. BioPlex 2200 Results for CDC T. gondii Reference Sera (n=100)

CDC Samples Reference Result BioPlex Pos (+) BioPlex Neg (–)
CMV IgG Pos (+) 66 0
Neg (–) 0 34


Note: Tested at Bio-Rad Laboratories

BioPlex™ 2200 ToRC IgG

Reagents, Calibrators, and Controls

ToRC IgG Reagent Pack

Each contains all the necessary reagents (sample diluent, beads and conjugate) to process up to 4 IgG results per sample.

ToRC IgG Calibrator Set

The ToRC IgG calibrators are multiplexed and ready-to-use.

ToRC IgG Control Set

The ToRC IgG controls are multiplexed and ready-to-use.

The BioPlex™ 2200 ToRC IgG kit is a multiplex flow immunoassay for the quantitative detection of IgG antibodies to Toxoplasma gondii and Rubella and the qualitative detection of IgG antibodies to Cytomegalovirus.

The system’s innovative eFlex™ software provides the ability to perform one, two or all three assays with add-on test capability without sample reruns.

BioPlex<sup>®</sup> 2200 ToRC IgG Reagent Pack
BioPlex® 2200 ToRC IgG Reagent Pack

665-1650
Reagents for 100 samples for IgG antibodies to Toxoplasma gondii, Rubella and Cytomegalovirus

List Price:   Inquire
Quantity:   Add to Quote
Add to Hot List  
BioPlex<sup>®</sup> 2200 ToRC IgG Calibrator Set
BioPlex® 2200 ToRC IgG Calibrator Set

663-1600
Set of multiplexed standards for calibration of results of ToRC screening based on IgG antibodies on BioPlex® 2200 system

List Price:   Inquire
Quantity:   Add to Quote
Add to Hot List  
BioPlex<sup>®</sup> 2200 ToRC IgG Control Set
BioPlex® 2200 ToRC IgG Control Set

663-1630
Set of multiplexed negative and multianalyte positive controls for ToRC IgG screening on the BioPlex® 2200 System

List Price:   Inquire
Quantity:   Add to Quote
Add to Hot List  
Number Description Options
T-157 BioPlex 2200 ToRC IgG Product Sheet Click to download [ Add to Cart (Free) ]
T-136 BioPlex 2200 Panel Menu Click to download [ Add to Cart (Free) ]
T-167 BioPlex 2200 Infectious Disease Panels Brochure Click to download [ Add to Cart (Free) ]