Model 1575 Immunowash™ Microplate Washer

Print

Model 1575 Immunowash™ Microplate Washer


Question:

How do I install a new manifold to my model 1575 Immunowash™ microplate washer?


Answer:

The washer tubing connects to the manifold using two metal connectors that sit on a square plastic housing at the end of the manifold. This plastic housing must be removed from the old manifold and transferred to the new one. Remove the single screw from the middle of the plastic housing to remove it from the existing manifold. Connect the plastic housing to the new manifold and screw them together. Put the new manifold into the cradle.

If the plastic housing is missing or damaged, it can be ordered through Technical Support by using part #100-1124. Note that this is not a standard catalog number.