Technique and Recommended Apparatus Gel or Tube Size,* Qty (W x L x Thickness) Typical Conditions** (Initial) Typical Conditions** (Final) Run Time PowerPac™ Power Supply
W V mA W V mA
Laemmli (SDS), O'Farrell Second Dimension (SDS)
PROTEAN® II xi Cell 160 x 160 x 1.5 mm, 2 gels 100 35(C) 350 35(C) 5 hr HV or Universal
PROTEAN II XL Cell 183 x 200 x 1.5 mm, 2 gels 100 35(C) 350 35(C) 5 hr HV or Universal
Criterion™ Cell 133 x 87 x 1.0 mm, 2 gels 200(C) 200 200(C) 80 55 min Basic or HC
Criterion™ 133 x 87 x 1.0 mm, 2 gels 300(C) 224 300(C) 164 20–26 min Basic, HV, or Universal
Mini-PROTEAN® Tetra Cell 83 x 73 x 1.0 mm, 4 gels 200(C) 240 200(C) 120 35–45 min Basic or HC
Mini-PROTEAN Tetra Cell 83 x 73 x 1.0 mm, 4 gels 300(C) 360 300(C) 228 15 min Basic, HV, or Universal
Mini-PROTEAN Tetra Cell 83 x 73 x 1.0 mm, 4 gels 400(C) 456 400(C) 416 10 min HV or Universal
High-Throughput Electrophoresis
PROTEAN Plus 200 x 205 x 1.0 mm, 12 gels 200(C) 1,000 200(C) 350 6 hr HC or Universal
Dodeca™ Cell 250 x 205 x 1.0 mm, 12 gels 200(C) 1,000 200(C) 350 6 hr HC or Universal
256 x 230 x 1.0 mm, 12 gels 150(C) 1,200 150(C) 300 18–20 hr HC or Universal
PROTEAN II xi/XL Multi-Cell 160 x 200 x 1.5 mm, 6 gels 150 480(C) 500 480(C) 5 hr Universal
Criterion™ Dodeca™ Cell 133 x 87 x 1.0 mm, 12 gels 200(C) 1,000–1,400 200(C) 400–500 55 min HC or Universal
Mini-PROTEAN® 3 Dodeca™ Cell 83 x 73 x 1.0 mm, 12 gels 200(C) 600 200(C) 360 45 min HC or Universal
IEF, O'Farrell First Dimension
PROTEAN II xi Cells 150 x 1.5 mm tubes, 4 (minimum) 800(C) 3.5 800(C) <1 16 hr HV or Universal
Mini-PROTEAN II Tube Cell 75 x 1.0 mm tubes, 8 (minimum) 750(C) 1 750(C) <1 3–4 hr HV or Universal
Preparative Electrofocusing
Rotofor® Cell 60 ml focusing chamber 15(C) 500 24 15(C) 1,200 10 4 hr HV
Mini Rotofor Cell 18 ml focusing chamber 12(C) 500 24 12(C) 1,200 10 4 hr HV
MicroRotofor™ Cell 2.5 ml focusing chamber 1(C) 100 10 1(C) 500 2 <3 hr HV or Universal
Preparative PAGE
Model 491 Prep Cell 10(C) 300 40 10(C) 400 30 3–8 hr HV or Universal
Mini Prep Cell 1(C) 200 5 1(C) 300 3 3–8 hr HV or Universal
Protein Electroelution
Model 422 Electro-Eluter 6 samples 200 60(C) 150 60(C) 3–4 hr Basic, HV, or Universal
Whole Gel Eluter 160 x 140 mm or larger 15 200(C) 15 200(C) 20 min Basic
Mini Whole Gel Eluter 55 x 65 mm or larger 10 100(C) 10 100(C) 20 min Basic
Polyacrylamide Analytical Electrofocusing
Model 111 Mini IEF Cell 125 x 65 x 0.4 mm 100(C) 6 100(C) 4 15 min HV or Universal
200(C) 6 200(C) 4 15 min Basic
450(C) 4 450(C) 1 1 hr Basic
DNA Restriction Analysis (Horizontal Mode)
Sub-Cell® GT Cell 150 x 200 x 5 mm 80(C) 55 80(C) 60 4 hr Basic or HC
Mini-Sub® Cell GT Cell 70 x 100 x 5 mm 50(C) 25 50(C) 30 2 hr Basic
Wide Mini-Sub Cell GT Cell 150 x 100 x 5 mm 50(C) 35 50(C) 40 2 hr Basic
DNA Sequencing
Sequi-Gen® GT System 380 x 500 x 0.4 mm 80(C) 1,850 30 80(C) 1,850 30 2–4 hr HV
SSCP
Sequi-Gen GT System 210 x 400 x 0.4 mm 40(C) 1,800 20 40(C) 1,800 20 2–3 hr HV
Microsatellite Mapping
Sequi-Gen GT System 210 x 400 x 0.4 mm 50(C) 2,100 25 50(C) 2,100 25 2–3 hr HV
Mutation Detection
DCode™ System 100 x 75 x 1.0 mm, 2 gels 130(C) 130(C) 2.5 hr Basic or HC
Western Blotting
Mini Trans-Blot® Cell 83 x 73 x 0.75 mm, 2 gels 100(C) 250 100(C) 450 1 hr HC
Criterion Blotter
 Wire electrodes 133 x 87 x 1.0 mm, 2 gels 100(C) 250 100(C) 450 1 hr HC
 Plate electrodes 133 x 87 x 1.0 mm, 2 gels 100(C) 650 100(C) 1,600 30 min HC
Trans-Blot® Cell
 Wire electrodes 200 x 160 x 1.5 mm, 1 gel 60(C) 210–250 60(C) 210—250 5 hr HC
 Plate electrodes 200 x 160 x 1.5 mm, 1 gel 100–150(C) 1,000–1,600 100–150(C) 1,000–1,600 1–5 hr HC
High-Intensity Transfer 200 x 160 x 1.5 mm, 1 gel 50–100(C) 1,600 50–100(C) 1,600 30 min HC
Trans-Blot Plus Cell 265 x 280 x 1.5 mm, 3 gels 100(C) 3,000 100(C) 3,000 30 min–1 hr HC
Semi-Dry Blotting
Trans-Blot SD Cell
 Protein 250 x 180 x 1.5 mm 15(C) 500 15(C) 200 15–30 min HC
 DNA/RNA 150 x 150 x 6 mm 15 650(C) 25 650(C) 10–30 min HC
 

* Sizes shown are typical for the corresponding apparatus. For running conditions for additional sizes, see the product instruction manuals.

** (C) = constant. Typical conditions are listed as guidelines only and can vary based on sample, buffers, etc.

PowerPac™ Power Supply Comparison
  PowerPac Basic PowerPac HC PowerPac Universal PowerPac HV
Output range (programmable)
  Volts 10–300 V 5–250 V 10–500 V 20–5,000 V
  Current 4–400 mA 0.01–3.0 A 0.01–2.5 A 0.01–500 mA
  Power 75 W (maximum) 1–300 W 1–500 W 1–400 W
Type of output (with automatic crossover) Constant voltage or constant current Constant voltage, constant current, or constant power Constant voltage, constant current, or constant power Constant voltage, constant current, constant power, or constant temperature
Timer 1–999 min 1 min–99 hr, 59 min 1 min–99 hr, 59 min 1 min–99 hr, 59 min
Volt-hour control No No Yes, 99,000 V-hr Yes, 99,000 V-hr
Pause/resume function Yes Yes Yes Yes
Display 3-digit LED 16-character x 2-line LCD 128 x 64 pixel, backlit 128 x 64 pixel, backlit
Programmable methods 1 method up to 3 steps, no storage capability Stores 9 methods, each with up to 9 steps Stores 9 methods, each with up to 9 steps
Real-time clock No No Yes Yes
Automatic recovery after power failure Yes Yes Yes Yes
Data transfer/archiving No No Yes, optional Yes
Temperature control No No No Via temperature probe; 30–90°C ± 2°C
Microampere readout No No No Yes
Safety features No-load detection; sudden load change detection, overload/short-circuit detection; overvoltage protection No-load detection; sudden load change detection; groundleak detection, overload/short-circuit detection; overvoltage protection No-load detection; sudden load change detection; groundleak detection, overload/short-circuit detection; overvoltage protection No-load detection; sudden load change detection; ground leak detection, arc detection;overload/short-circuit detection; overvoltage protection
Operating conditions 0–40°C; 0–95% humidity 0–40°C; 0–95% humidity 0–40°C; 0–95% humidity 0–40°C; 0–95% humidity
Stackable Yes Yes Yes Yes
Number of output jacks 4 sets in parallel 4 sets in parallel 4 sets in parallel 4 sets in parallel
Regulatory EN-61010, CE EN-61010, CE EN-61010, CE EN-61010, CE
IQ/OQ protocols No No Yes, optional Yes, optional
Input power (actual) 90–120 or 198–264 VAC, 50/60 Hz, autoswitching 90–120 or 198–264 VAC, 50/60 Hz, autoswitching 90–120 or 198–264 VAC,50/60 Hz, autoswitching 90–120 or 198–264 VAC,50/60 Hz, autoswitching
Dimensions (W x D x H) 21 x 24.5 x 6.5 cm 25 x 28.5 x 8 cm 27.5 x 34 x 10 cm 27.5 x 34 x 10 cm
Weight 1.1 kg (2.4 lb) 2.0 kg (4.4 lb) 2.5 kg (5.5 lb) 2.85 kg (6.3 lb)
Scroll upScroll down


The PowerPac™ Basic power supply is recommended for basic applications such as submerged horizontal and mini vertical gel electrophoresis.

Features and Benefits

  • Programming in a compact, stackable case
  • Timer control and constant voltage or constant current output
  • Ability to pause and resume a run

Applications and Uses

  • Nucleic acid gel electrophoresis
  • Protein gel electrophoresis

To select the power supply that best fits your needs, see the Power Supply Selection Guide or the PowerPac Power Supply Comparison chart.

 

Output range (programmable) 10–300 V, fully adjustable in 1 V steps
4–400 mA, fully adjustable in 1 mA steps
75 W (maximum)
Type of output Constant voltage or constant current with automatic crossover
Output terminals 4 pair recessed banana jacks in parallel
Timer 1 min–99 hr 59 min, fully adjustable
Pause/resume function Yes
Display 3-digit LED
Operating conditions 0–40°C, 0–90% humidity in the absence of condensation
Regulatory EN-61010, CE
Safety features No-load detection, sudden load change detection, overload/short-circuit detection, overvoltage protection
Input protection Fuse on both hot and neutral
Input power (actual) 90–120 or 198–264 VAC, 50/60 Hz, autoswitching
Dimensions (W x L x H), cm/in 21 x 24.5 x 6.5/8.3 x 9.6 x 2.6
Weight, kg/lb 1.1/2.4
x

Account price failed to display. Please try reloading the page.Reload Page

If you are placing an order, you may proceed with your order; the account price will be applied if it is lower than the
list price. If the problem persists, please call customer service at or send an
email to

PowerPac™ Basic Power Supply

1645050
100–120/220–240 V, power supply for basic applications such as submerged horizontal gel electrophoresis, includes power cord

0 1645050 /templatedata/internet/product_SKU/data/LSR/Electrophoresis-and-Blotting/Power-Supplies/164-5050_0921120318989.xml 1645050 /default/main/bio-rad/WORKAREA/default 09/21/09 12:03 PM en_US jebrbibc 2018-02-28 05:02:10 true true internet/product_SKU jebrbibc templatedata/internet/product_SKU/data/LSR/Electrophoresis-and-Blotting/Power-Supplies/164-5050_0921120318989.xml LSR 2018-03-06 20:27:42 /default/main/bio-rad/STAGING N /default/main/bio-rad en 09/21/19 12:03 PM AE,AI,AL,AM,AR,AT,AU,AZ,BA,BD,BE,BF,BG,BH,BN,BO,BR,BW,CA,CH,CL,CM,CN,CO,CR,CY,CZ,DE,DK,DO,DZ,EC,EE,EG,EH,ER,ES,ET,FI,FM,FO,FR,GA,GE,GF,GH,GP,GR,GT,GU,HK,HN,HR,HT,HU,ID,IE,IL,IN,IS,IT,JM,JO,JP,KE,KH,KR,KW,KZ,LB,LI,LK,LT,LU,LV,MA,MD,MG,MK,ML,MO,MQ,MS,MT,MU,MX,MY,NG,NI,NL,NO,NP,NZ,OM,PA,PE,PF,PG,PH,PK,PL,PR,PS,PT,PW,PY,QA,RO,RS,RU,SA,SB,SE,SG,SI,SK,SN,ST,SV,TG,TH,TN,TO,TR,TT,TW,TZ,UA,UG,UK,US,UY,UZ,VA,VE,VU,XK,YE,ZA 2018-03-06 20:27:42 /default/main/bio-rad/WORKAREA/default/templatedata/internet/product_SKU/data/LSR/Electrophoresis-and-Blotting/Power-Supplies/164-5050_0921120318989.xml templatedata/internet/product_SKU/data/LSR/Electrophoresis-and-Blotting/Power-Supplies/164-5050_0921120318989.xml 1645050-powerpac-basic-power-supply 1645050 PowerPac Basic Power Supply PowerPac™ Basic Power Supply 100–120/220–240 V, power supply for basic applications such as submerged horizontal gel electrophoresis, includes power cord 100–120/220–240 V, power supply for basic applications such as submerged horizontal gel electrophoresis, includes power cord /webroot/web/images/lsr/products/electrophoresis/sku_thumb/global/164-5050_tn.jpg /webroot/web/images/lsr/products/electrophoresis/sku_view/global/164-5050_view.jpg /webroot/web/images/lsr/products/electrophoresis/sku_feature/global/164-5050_feature.jpg PowerPac Basic Power Supply #164-5050 RepEx PowerPac Basic Power Supply #164-5050 100–120/220–240 V, power supply for basic applications such as submerged horizontal gel electrophoresis, includes power cord 1645050, powerpac, basic, power, supply, electrophoresis, gel, protein, dna, trans-blot, blot, transfer 09/21/09 12:03 PM 09/21/19 12:03 PM AE,AI,AL,AM,AR,AT,AU,AZ,BA,BD,BE,BF,BG,BH,BN,BO,BR,BW,CA,CH,CL,CM,CN,CO,CR,CY,CZ,DE,DK,DO,DZ,EC,EE,EG,EH,ER,ES,ET,FI,FM,FO,FR,GA,GE,GF,GH,GP,GR,GT,GU,HK,HN,HR,HT,HU,ID,IE,IL,IN,IS,IT,JM,JO,JP,KE,KH,KR,KW,KZ,LB,LI,LK,LT,LU,LV,MA,MD,MG,MK,ML,MO,MQ,MS,MT,MU,MX,MY,NG,NI,NL,NO,NP,NZ,OM,PA,PE,PF,PG,PH,PK,PL,PR,PS,PT,PW,PY,QA,RO,RS,RU,SA,SB,SE,SG,SI,SK,SN,ST,SV,TG,TH,TN,TO,TR,TT,TW,TZ,UA,UG,UK,US,UY,UZ,VA,VE,VU,XK,YE,ZA en LSR /LSR/Electrophoresis-and-Blotting/Power-Supplies N 0 <p>Use the PowerPac Basic Power Supply for applications such as mini and midi vertical gel and submerged horizontal electrophoresis, and blotting using the <a href="/en-us/product/mini-trans-blot-cell?ID=589ca8f7-5751-487a-a453-571ee8cc8b7e">Mini Trans-Blot<sup>&reg;</sup> Cell</a>. The power supply is fully programmable for either constant current or voltage, and either timed or untimed.</p> <h2 class="descHeader">Features include:</h2> <ul> <li>Pause mode for editing run parameters, or stopping and resuming the run</li> <li>Power failure detection in timed mode; when power returns run will complete</li> <li>Automatic detection of rapid changes in resistance, and no load conditions</li> <li>Compact and stackable</li> <li>Flip-down legs to change viewing angle </li> </ul> <p>To prevent damage to electrophoresis cells, the PowerPac Basic Power Supply has automatic crossover. When the set limit of the nonconstant parameter (either voltage or current, depending on settings) is reached, and the power capability of the unit is not exceeded, the power supply will make the nonconstant parameter the new constant parameter.</p> <p>In addition to the <a href="/en-us/product/powerpac-basic-power-supply?ID=bea5dea1-cef0-43ad-8af5-b2c0287f6e07">PowerPac Basic Power Supply</a>, other power supplies include:</p> <ul> <li><a href="/en-us/product/powerpac-universal-power-supply">PowerPac Universal Power Supply</a> &mdash; for a broad range of applications</li> <li><a href="/en-us/product/powerpac-hc-high-current-power-supply">PowerPac HC High-Current Power Supply</a> &mdash; for high current applications including western blotting</li> <li><a href="/en-us/product/powerpac-hv-high-voltage-power-supply">PowerPac HV High Voltage Power Supply</a> &mdash; for high voltage applications including IEF and DNA sequencing</li> </ul> <h2 class="descHeader">Related Products</h2> <ul> <li><a href="/en-us/category/protein-gels">Protein Gels</a></li> <li><a href="/en-us/product/mini-protean-tetra-vertical-electrophoresis-cell?ID=N3F2UD4VY">Mini-PROTEAN<sup>&reg;</sup> Tetra Electrophoresis Cells</a></li> </ul> <ul> </ul> <h2 class="descHeader">Related Categories</h2> <ul> <li><a href="/en-us/category/western-blotting?ID=9324fd3c-6af4-4551-831e-8db6fa3f3452">Western Blotting</a></li> <li><a href="/en-us/category/protein-electrophoresis-blotting">Protein Electrophoresis and Blotting</a></li> <li><a href="/en-us/category/nucleic-acid-electrophoresis-blotting">Nucleic Acid Electrophoresis and Blotting</a></li> </ul> <h2 class="descHeader">Supporting Documents and Resources</h2> <ul> <li>PowerPac Basic Power Supply <a href="/webroot/web/pdf/lsr/literature/4006213.pdf">Instruction Manual</a></li> <li> <a href="/webroot/web/pdf/lsr/literature/Bulletin_6040.pdf">Protein Electrophoresis Guide</a></li> </ul> <ul style="padding-left: 30px;"> </ul> <h2 class="descHeader">Related Applications</h2> <ul> <li><a href="/en-us/applications-technologies/introduction-protein-electrophoresis?ID=LUSOVO47B">Protein Electrophoresis</a></li> <li><a href="/en-us/applications-technologies/introduction-western-blotting?ID=LUSPPAKG4">Western Blotting</a></li> </ul> <ul style="padding-left: 30px;"> </ul> /webroot/web/html/lsr/products/SKUs/autopopulated/ 164-5050_0921120318989_en.html /webroot/web/html/lsr/products/SKUs/autopopulated//164-5050_0921120318989_en.html Output specifications 1938 10&ndash;300 V, fully adjustable in 1 V steps<br> 4&ndash;400 mA, fully adjustable in 1 mA steps<br> 75 W (maximum) 1 0 Type of output 2284 Constant voltage or constant current with automatic crossover 1 0 Output terminals 1939 4 pair recessed banana jacks in parallel 1 0 Timer control 2263 1 min&ndash;99 hr 59 min, fully adjustable 1 0 Pause/resume function 1950 Yes 1 0 Display 1573 3-digit LED 1 0 Operating conditions 1897 0&ndash;40&deg;C; 0&ndash;95% humidity in absence of condensation 1 0 Safety compliance 2100 EN61010 1 0 Safety features 2101 No-load detection; rapid resistance change detection, ground leak detection, overload/short circuit detection, overvoltage protection, over-temperature protection 1 0 Input protection 1737 Fuse on hot and neutral 1 0 Dimensions (W x D x H), cm/in 1562 21 x 24.5 x 6.5/8.3 x 9.6 x 2.6 1 0 Weight, kg/lb 2338 1.1/2.4 1 0 false false false /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645050&CartQuote=QT&quantity= false true false CZK13,720.00 false null /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645050&AddToDefaultWishList=true&quantity= true false true true false true false true /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645050&quantity= 1645050
List Price: 
 
CZK13,720.00 
LoadingLoading...
EDU Price:
  What is this?
Your Price: 
 
LoadingLoading...
Your Price:
 

Communication error

Reload Page

Quantity:   Add to Cart
 
 
x

Account price failed to display. Please try reloading the page.Reload Page

If you are placing an order, you may proceed with your order; the account price will be applied if it is lower than the
list price. If the problem persists, please call customer service at or send an
email to

PowerPac Adaptor

1645062
Pkg of 1, 2 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies

2 1645062 /templatedata/internet/product_SKU/data/LSR/Electrophoresis-and-Blotting/Power-Supplies/164-5062_0921120317845.xml 1645062 09/21/09 12:03 PM N en 09/21/19 12:03 PM true true internet/product_SKU AE,AI,AL,AM,AR,AT,AU,AZ,BA,BD,BE,BF,BG,BH,BN,BO,BR,BW,CA,CH,CL,CM,CN,CO,CR,CY,CZ,DE,DK,DO,DZ,EC,EE,EG,EH,ER,ES,ET,FI,FM,FO,FR,GA,GE,GF,GH,GP,GR,GT,GU,HK,HN,HR,HT,HU,ID,IE,IL,IN,IS,IT,JM,JO,JP,KE,KH,KR,KW,KZ,LB,LI,LK,LT,LU,LV,MA,MD,MG,MK,ML,MO,MQ,MS,MT,MU,MX,MY,NG,NI,NL,NO,NP,NZ,OM,PA,PE,PF,PG,PH,PK,PL,PR,PS,PT,PW,PY,QA,RO,RS,RU,SA,SB,SE,SG,SI,SK,SN,ST,SV,TG,TH,TN,TO,TR,TT,TW,TZ,UA,UG,UK,US,UY,UZ,VA,VE,VU,XK,YE,ZA LSR 2017-11-14 07:08:18 1645062-powerpac-adaptor 1645062 PowerPac Adaptor PowerPac Adaptor Pkg of 1, 2 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies Pkg of 1, 2 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies Pkg of 1, 2 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies /webroot/web/images/lse/products/equipment_supplies/sku_thumb/global/164-5062_tn.jpg /webroot/web/images/lse/products/equipment_supplies/sku_view/global/164-5062_view.jpg /webroot/web/images/lse/products/equipment_supplies/sku_feature/global/164-5062_feature.jpg RepEx 164-5062 Pkg of 1, 2 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies 09/21/09 12:03 PM 09/21/19 12:03 PM AE,AI,AL,AM,AR,AT,AU,AZ,BA,BD,BE,BF,BG,BH,BN,BO,BR,BW,CA,CH,CL,CM,CN,CO,CR,CY,CZ,DE,DK,DO,DZ,EC,EE,EG,EH,ER,ES,ET,FI,FM,FO,FR,GA,GE,GF,GH,GP,GR,GT,GU,HK,HN,HR,HT,HU,ID,IE,IL,IN,IS,IT,JM,JO,JP,KE,KH,KR,KW,KZ,LB,LI,LK,LT,LU,LV,MA,MD,MG,MK,ML,MO,MQ,MS,MT,MU,MX,MY,NG,NI,NL,NO,NP,NZ,OM,PA,PE,PF,PG,PH,PK,PL,PR,PS,PT,PW,PY,QA,RO,RS,RU,SA,SB,SE,SG,SI,SK,SN,ST,SV,TG,TH,TN,TO,TR,TT,TW,TZ,UA,UG,UK,US,UY,UZ,VA,VE,VU,XK,YE,ZA en LSR /LSR/Electrophoresis-and-Blotting/Power-Supplies N 0 false false false /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645062&CartQuote=QT&quantity= false true false CZK1,290.00 false null /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645062&AddToDefaultWishList=true&quantity= true false true true false true false true /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645062&quantity= 1645062
List Price: 
 
CZK1,290.00 
LoadingLoading...
EDU Price:
  What is this?
Your Price: 
 
LoadingLoading...
Your Price:
 

Communication error

Reload Page

Quantity:   Add to Cart
 
 
PowerPac Adaptor

1645064
Pkg of 1, 4 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies

3 1645064 /templatedata/internet/product_SKU/data/LSR/Electrophoresis-and-Blotting/Power-Supplies/164-5064_0921120317864.xml 1645064 09/21/09 12:03 PM N en 09/21/19 12:03 PM true true internet/product_SKU AE,AI,AL,AM,AR,AT,AU,AZ,BA,BD,BE,BF,BG,BH,BN,BO,BR,BW,CA,CH,CL,CM,CN,CO,CR,CY,CZ,DE,DK,DO,DZ,EC,EE,EG,EH,ER,ES,ET,FI,FM,FO,FR,GA,GE,GF,GH,GP,GR,GT,GU,HK,HN,HR,HT,HU,ID,IE,IL,IN,IS,IT,JM,JO,JP,KE,KH,KR,KW,KZ,LB,LI,LK,LT,LU,LV,MA,MD,MG,MK,ML,MO,MQ,MS,MT,MU,MX,MY,NG,NI,NL,NO,NP,NZ,OM,PA,PE,PF,PG,PH,PK,PL,PR,PS,PT,PW,PY,QA,RO,RS,RU,SA,SB,SE,SG,SI,SK,SN,ST,SV,TG,TH,TN,TO,TR,TT,TW,TZ,UA,UG,UK,US,UY,UZ,VA,VE,VU,XK,YE,ZA LSR 2017-11-14 07:08:18 1645064-powerpac-adaptor 1645064 PowerPac Adaptor PowerPac Adaptor Pkg of 1, 4 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies Pkg of 1, 4 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies Pkg of 1, 4 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies /webroot/web/images/lse/products/equipment_supplies/sku_thumb/global/164-5064_tn.jpg /webroot/web/images/lse/products/equipment_supplies/sku_view/global/164-5064_view.jpg /webroot/web/images/lse/products/equipment_supplies/sku_feature/global/164-5064_feature.jpg RepEx 164-5064 Pkg of 1, 4 mm, adaptor, for converting non-IEC-certified cells to fit deeper output terminals of PowerPac power supplies 164-5064, 1645064, powerpac, adaptor, power, supply, supplies, accessories, electrophoresis 09/21/09 12:03 PM 09/21/19 12:03 PM AE,AI,AL,AM,AR,AT,AU,AZ,BA,BD,BE,BF,BG,BH,BN,BO,BR,BW,CA,CH,CL,CM,CN,CO,CR,CY,CZ,DE,DK,DO,DZ,EC,EE,EG,EH,ER,ES,ET,FI,FM,FO,FR,GA,GE,GF,GH,GP,GR,GT,GU,HK,HN,HR,HT,HU,ID,IE,IL,IN,IS,IT,JM,JO,JP,KE,KH,KR,KW,KZ,LB,LI,LK,LT,LU,LV,MA,MD,MG,MK,ML,MO,MQ,MS,MT,MU,MX,MY,NG,NI,NL,NO,NP,NZ,OM,PA,PE,PF,PG,PH,PK,PL,PR,PS,PT,PW,PY,QA,RO,RS,RU,SA,SB,SE,SG,SI,SK,SN,ST,SV,TG,TH,TN,TO,TR,TT,TW,TZ,UA,UG,UK,US,UY,UZ,VA,VE,VU,XK,YE,ZA en LSR /LSR/Electrophoresis-and-Blotting/Power-Supplies N 0 false false false /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645064&CartQuote=QT&quantity= false true false CZK1,290.00 false null /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645064&AddToDefaultWishList=true&quantity= true false true true false true false true /prd/en/CZ/adirect/biorad?cmd=ajaxAddToCart&AddToCartProductID=1645064&quantity= 1645064
List Price: 
 
CZK1,290.00 
LoadingLoading...
EDU Price:
  What is this?
Your Price: 
 
LoadingLoading...
Your Price:
 

Communication error

Reload Page

Quantity:   Add to Cart
 
 
Number Description Options
6371 Electrophoresis Power Supplies Brochure, Rev C Click to download [ Add to Cart (Free) ]
4006213 Instruction Manual, PowerPac Basic Power Supply, Rev F Click to download

This product contains electrical or electronic material. Learn more about Bio-Rad's EU Recycle Program

Get Help

Call us at 00 800 00 BIORAD (00 800 00 24 67 23) or
+36 1 459 6192, +420 241 431 660

Sign Up for Bio-Rad Updates!
Enter your email address below to receive your choice of the latest news, promotions, and more.
To give you the very best experience, the cookie settings on Bio-Rad.com are set to "allow all cookies". To change your settings please click  here.  Close