Technique and Recommended Apparatus Gel or Tube Size,* Qty (W x L x Thickness) Typical Conditions** (Initial) Typical Conditions** (Final) Run Time PowerPac™ Power Supply
W V mA W V mA
Laemmli (SDS), O'Farrell Second Dimension (SDS)
PROTEAN® II xi cell 160 x 160 x 1.5 mm, 2 gels 100 35(C) 350 35(C) 5 hr HV or Universal
PROTEAN II XL cell 183 x 200 x 1.5 mm, 2 gels 100 35(C) 350 35(C) 5 hr HV or Universal
Criterion™ cell 133 x 87 x 1.0 mm, 2 gels 200(C) 200 200(C) 80 55 min Basic or HC
Criterion™ 133 x 87 x 1.0 mm, 2 gels 300(C) 224 300(C) 164 20–26 min Basic, HV, or Universal
Mini-PROTEAN® Tetra cell 83 x 73 x 1.0 mm, 4 gels 200(C) 240 200(C) 120 35–45 min Basic or HC
Mini-PROTEAN® Tetra Cell 83 x 73 x 1.0 mm, 4 gels 300(C) 360 300(C) 228 15 min Basic, HV, or Universal
Mini-PROTEAN® Tetra Cell 83 x 73 x 1.0 mm, 4 gels 400(C) 456 400(C) 416 10 min HV or Universal
High-Throughput Electrophoresis
PROTEAN Plus 200 x 205 x 1.0 mm, 12 gels 200(C) 1,000 200(C) 350 6 hr HC or Universal
Dodeca™ cell 250 x 205 x 1.0 mm, 12 gels 200(C) 1,000 200(C) 350 6 hr HC or Universal
256 x 230 x 1.0 mm, 12 gels 150(C) 1,200 150(C) 300 18–20 hr HC or Universal
PROTEAN II xi/XL multi-cell 160 x 200 x 1.5 mm, 6 gels 150 480(C) 500 480(C) 5 hr Universal
Criterion™ Dodeca™ cell 133 x 87 x 1.0 mm, 12 gels 200(C) 1,000–1,400 200(C) 400–500 55 min HC or Universal
Mini-PROTEAN® 3 Dodeca™ cell 83 x 73 x 1.0 mm, 12 gels 200(C) 600 200(C) 360 45 min HC or Universal
IEF, O'Farrell First Dimension
PROTEAN II xi cells 150 x 1.5 mm tubes, 4 (minimum) 800(C) 3.5 800(C) <1 16 hr HV or Universal
Mini-PROTEAN II tube cell 75 x 1.0 mm tubes, 8 (minimum) 750(C) 1 750(C) <1 3–4 hr HV or Universal
Preparative Electrofocusing
Rotofor® cell 60 ml focusing chamber 15(C) 500 24 15(C) 1,200 10 4 hr HV
Mini Rotofor cell 18 ml focusing chamber 12(C) 500 24 12(C) 1,200 10 4 hr HV
MicroRotofor™ cell 2.5 ml focusing chamber 1(C) 100 10 1(C) 500 2 <3 hr HV or Universal
Preparative PAGE
Model 491 prep cell 10(C) 300 40 10(C) 400 30 3–8 hr HV or Universal
Mini prep cell 1(C) 200 5 1(C) 300 3 3–8 hr HV or Universal
Protein Electroelution
Model 422 electro-eluter 6 samples 200 60(C) 150 60(C) 3–4 hr Basic, HV, or Universal
Whole gel eluter 160 x 140 mm or larger 15 200(C) 15 200(C) 20 min Basic
Mini whole gel eluter 55 x 65 mm or larger 10 100(C) 10 100(C) 20 min Basic
Polyacrylamide Analytical Electrofocusing
Model 111 mini IEF cell 125 x 65 x 0.4 mm 100(C) 6 100(C) 4 15 min HV or Universal
200(C) 6 200(C) 4 15 min Basic
450(C) 4 450(C) 1 1 hr Basic
DNA Restriction Analysis (Horizontal Mode)
Sub-Cell® GT cell 150 x 200 x 5 mm 80(C) 55 80(C) 60 4 hr Basic or HC
Mini-Sub® cell GT cell 70 x 100 x 5 mm 50(C) 25 50(C) 30 2 hr Basic
Wide Mini-Sub cell GT cell 150 x 100 x 5 mm 50(C) 35 50(C) 40 2 hr Basic
DNA Sequencing
Sequi-Gen® GT system 380 x 500 x 0.4 mm 80(C) 1,850 30 80(C) 1,850 30 2–4 hr HV
SSCP
Sequi-Gen GT system 210 x 400 x 0.4 mm 40(C) 1,800 20 40(C) 1,800 20 2–3 hr HV
Microsatellite Mapping
Sequi-Gen GT system 210 x 400 x 0.4 mm 50(C) 2,100 25 50(C) 2,100 25 2–3 hr HV
Mutation Detection
DCode™ system 100 x 75 x 1.0 mm, 2 gels 130(C) 130(C) 2.5 hr Basic or HC
Western Blotting
Mini Trans-Blot® cell 83 x 73 x 0.75 mm, 2 gels 100(C) 250 100(C) 450 1 hr HC
Criterion blotter
  Wire electrodes 133 x 87 x 1.0 mm, 2 gels 100(C) 250 100(C) 450 1 hr HC
  Plate electrodes 133 x 87 x 1.0 mm, 2 gels 100(C) 650 100(C) 1,600 30 min HC
Trans-Blot® cell
  Wire electrodes 200 x 160 x 1.5 mm, 1 gel 60(C) 210–250 60(C) 210—250 5 hr HC
  Plate electrodes 200 x 160 x 1.5 mm, 1 gel 100–150(C) 1,000–1,600 100–150(C) 1,000–1,600 1–5 hr HC
  High-intensity transfer 200 x 160 x 1.5 mm, 1 gel 50–100(C) 1,600 50–100(C) 1,600 30 min HC
Trans-Blot Plus cell 265 x 280 x 1.5 mm, 3 gels 100(C) 3,000 100(C) 3,000 30 min–1 hr HC
Semi-Dry Blotting
Trans-Blot SD cell
  Protein 250 x 180 x 1.5 mm 15(C) 500 15(C) 200 15–30 min HC
  DNA/RNA 150 x 150 x 6 mm 15 650(C) 25 650(C) 10–30 min HC
 

* Sizes shown are typical for the corresponding apparatus. For running conditions for additional sizes, see the product instruction manuals.

** (C) = constant. Typical conditions are listed as guidelines only and can vary based on sample, buffers, etc.

PowerPac™ Power Supply Comparison
  PowerPac Basic PowerPac HC PowerPac Universal PowerPac HV
Output range (programmable)
  Volts 10–300 V 5–250 V 10–500 V 20–5,000 V
  Current 4–400 mA 0.01–3.0 A 0.01–2.5 A 0.01–500 mA
  Power 75 W (maximum) 1–300 W 1–500 W 1–400 W
Type of output (with automatic crossover) Constant voltage or constant current Constant voltage, constant current, or constant power Constant voltage, constant current, or constant power Constant voltage, constant current, constant power, or constant temperature
Timer 1–999 min 1 min–99 hr, 59 min 1 min–99 hr, 59 min 1 min–99 hr, 59 min
Volt-hour control No No Yes, 99,000 V-hr Yes, 99,000 V-hr
Pause/resume function Yes Yes Yes Yes
Display 3-digit LED 16-character x 2-line LCD 128 x 64 pixel, backlit 128 x 64 pixel, backlit
Programmable methods 1 method up to 3 steps, no storage capability Stores 9 methods, each with up to 9 steps Stores 9 methods, each with up to 9 steps
Real-time clock No No Yes Yes
Automatic recovery after power failure Yes Yes Yes Yes
Data transfer/archiving No No Yes, optional Yes
Temperature control No No No Via temperature probe; 30–90°C ± 2°C
Microampere readout No No No Yes
Safety features No-load detection; sudden load change detection, overload/short-circuit detection; overvoltage protection No-load detection; sudden load change detection; groundleak detection, overload/short-circuit detection; overvoltage protection No-load detection; sudden load change detection; groundleak detection, overload/short-circuit detection; overvoltage protection No-load detection; sudden load change detection; ground leak detection, arc detection;overload/short-circuit detection; overvoltage protection
Operating conditions 0–40°C; 0–95% humidity 0–40°C; 0–95% humidity 0–40°C; 0–95% humidity 0–40°C; 0–95% humidity
Stackable Yes Yes Yes Yes
Number of output jacks 4 sets in parallel 4 sets in parallel 4 sets in parallel 4 sets in parallel
Regulatory EN-61010, CE EN-61010, CE EN-61010, CE EN-61010, CE
IQ/OQ protocols No No Yes, optional Yes, optional
Input power (actual) 90–120 or 198–264 VAC, 50/60 Hz, autoswitching 90–120 or 198–264 VAC, 50/60 Hz, autoswitching 90–120 or 198–264 VAC,50/60 Hz, autoswitching 90–120 or 198–264 VAC,50/60 Hz, autoswitching
Dimensions (W x D x H) 21 x 24.5 x 6.5 cm 25 x 28.5 x 8 cm 27.5 x 34 x 10 cm 27.5 x 34 x 10 cm
Weight 1.1 kg (2.4 lb) 2.0 kg (4.4 lb) 2.5 kg (5.5 lb) 2.85 kg (6.3 lb)

The PowerPac™ HC power supply is recommended for high-current applications; it has an output of 250 V, 3.0 A, and 300 W. The PowerPac HC power supply is the perfect choice for western blotting using any of the Bio-Rad blotting cells.

Features and Benefits

  • Ideal for high-throughput electrophoresis
  • Two-line, 16-character LCD allows programming from a single screen and monitoring of run progress at a glance by displaying all parameters at once
  • Intuitive programming permits multistep methods and timed runs with choice of constant voltage, constant current, or constant power output

Applications and Uses

  • Western blotting (tank or semi-dry transfers)
  • Nucleic acid gel electrophoresis
  • Protein gel electrophoresis

To select the power supply that best fits your needs, see the Power Supply Selection Guide or the PowerPac Power Supply Comparison chart.

 

Output specifications 250 V, 3.0 A, 300 W
Output range (programmable) 5–250 V, fully adjustable in 1 V increments
0.01–3.0 A, fully adjustable in 0.01 A increments
1–300 W, fully adjustable in 1 W increments
Type of output Constant voltage, current, or power with automatic crossover
Output terminals 4 pair recessed banana jacks in parallel
Timer 1 min–99 hr 59 min, fully adjustable
Pause/resume function Yes
Display 2-line, 16-character backlit LCD
Operating conditions 0–40°C, 0–90% humidity in the absence of condensation
Safety compliance EN61010
EMI Conforms to CE standards for emissions and immunity class A, tested only at 230 V, TÜV EMC certification
Safety features No-load detection, rapid resistance change detection, ground leak detection, overload/short circuit detection, overvoltage protection, overtemperature protection
Input protection Fuse on hot and neutral
Input power (nominal) 100–120/220–240 VAC, autoswitching
Stackable Yes
Dimensions (W x L x H), cm/in 25 x 28.5 x 8/9.8 x 11.2 x 3.1
Weight, kg/lb 2.0/4.4
PowerPac HC Power Supply #164-5052
PowerPac™ HC Power Supply

164-5052
100–120/220–240 V, power supply for high-current applications such as western blotting, includes power cord